REVIEW OF SCIENTIFIC INSTRUMENTS
卷 71, 期 9, 页码 3274-3289出版社
AMER INST PHYSICS
DOI: 10.1063/1.1287637
关键词
-
A method for reducing time sequences of raw scattering images to intensity time-autocorrelation functions is presented. The procedure is based on the use of a charge coupled device (CCD) area detector, and optimized for operating in the regime of short data batches. Its application to x-ray photon correlation spectroscopy (XPCS) measurements is described in detail. Using a slow-scan CCD, we explain how to achieve data acquisition on a 30 ms or faster time scale, while simultaneously acquiring data from many coherence areas in parallel. The statistical uncertainties of the acquired XPCS data are quantified experimentally, and compared to the theoretically expected noise levels of the correlation functions. (C) 2000 American Institute of Physics. [S0034-6748(00)03509-7].