期刊
REVIEW OF SCIENTIFIC INSTRUMENTS
卷 71, 期 9, 页码 3254-3259出版社
AMER INST PHYSICS
DOI: 10.1063/1.1287626
关键词
-
We report on the excellent performance of a newly constructed soft x-ray helical undulator beamline BL25SU of SPring-8 for photon energies 500-1800 eV. The full beamline was designed to perform very high resolution soft x-ray spectroscopy of solids with using high brilliance, highly circularly polarized undulator radiation. The grazing incidence monochromator employs varied-line-spacing plane gratings which operate in convergent light from a spherical mirror and focuses monochromatic light onto the exit slit. A resolving power in excess of 15 000 was measured at 540 and 870 eV for a grating with a central groove density of 600 lines/mm from the photoemission spectra of Au. A resolving power of more than 20 000 is estimated near 870 eV for a grating with a central groove density of 1000 lines/mm. A photon flux of more than 1x10(11) photons/s/100 mA/0.02% b.w. is supplied onto the sample between 500 and 1800 eV with very low amount of higher-order light. The low heat load from the twin-helical undulator gives high stability to all optical components, which is essential to obtain high resolution in a wide energy region. Three experimental stations are installed in tandem on this beamline for various high resolution spectroscopy measurements. (C) 2000 American Institute of Physics. [S0034- 6748(00)02409-6].
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐