IEEE TRANSACTIONS ON MAGNETICS
卷 36, 期 5, 页码 3649-3651出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/20.908929
关键词
3-D magnetic field analysis; biomagnetic measurements; combination of magnetic and conductive layers; magnetically shielded room; shielding factor
This paper describes the optimal design of magnetically shielded room used for measurements of extremely weak magnetic fields such as biomagnetic measurements, in which the higher shielding factor is required in the frequency range from 0.01 Hz to several ten Hz at a low cost, The 3-D magnetic field analysis is carried out by using the finite element method taking account of eddy currents to estimate an effective combination of magnetic and conductive layers for the design of wall and door. It is confirmed that as regards the wall design, an Al layer itself can increase the shielding effect by a sandwiched PC-AL-PC (PC: permalloy, Al: aluminum) combination even at frequencies at which only Al layer has no shielding effect, When the wall is composed by four lavers [PC (inner)-PC (middle)-Al-PC (outer)], the door composed by the Al and outer PC layers is supposed to be optimal from the standpoints of shielding effect and cost. Its shielding factor is comparable to that of the door composed by Al, inner and middle PC's.