4.6 Article

Multi-element analysis by inductively coupled plasma optical emission spectrometry of airborne particulate matter collected with a low-pressure cascade impactor

期刊

ANALYST
卷 125, 期 10, 页码 1855-1859

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/b003048l

关键词

-

向作者/读者索取更多资源

A method was developed for the elemental analysis of size segregated particles ranging from 0.03 to 10 mu m. Sampling and analysis problems are discussed in this paper. Particles were collected with a Dekati low-pressure cascade impactor. PTFE filters coated with oleic acid were used as substrate. Particles were microwave digested in closed vessels. The optimum digestion mixture was composed of HNO3 (1 mL), HF (50 mu L) and H2O (1 mL). The optimal power setting and digestion time were studied in order to achieve an efficient digestion. A ca. 35 min microwave digestion cycle at a 650 W maximum power allowed complete digestion of the samples. Special emphasis was placed on the pressure in the closed vessels to avoid sample losses. Solution samples were analysed by inductively coupled plasma optical emission spectrometry using an ultrasonic nebuliser for 18 elements (Al, Ba, Ca, Cd, Cr, Cu, Fe, K, Mg, Mn, Na, Ni, Pb, S, Sr, Ti, V, Zn). This procedure was tested with NIST Standard Reference Material 1648 Urban Particulate. Recoveries for certified elements ranged from 95 to 105% except for Al (90%). The influence of cascade impactor materials was investigated with 44 field samples. Strong artefacts due to contamination were shown for analysis at environmental concentrations of Al, Cr, Mn and Ni.