期刊
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL
卷 47, 期 2, 页码 361-368出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/58.827421
关键词
-
The frequency flicker of quartz resonators can be derived from the measurement of S phi(f), i.e., the power spectrum density of phase fluctuations phi. The interferometric method appears to be the best choice to measure the phase fluctuations of the quartz, resonators because of its high sensitivity in the low power conditions, which is required for this type of resonator. Combining these two ideas, we built an instrument suitable to measure the frequency flicker floor of the quartz resonators, and we measured the stability of some 10-MHz high performance resonators as a function of the dissipated power. The stability limit of our instrument, described in terms of Allan deviation sigma(y)(tau), is of some 10(-14).
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐