A high voltage, microsecond regime, pulsed glow discharge has been applied to a standard Grimm type source on a LECO SA-2000 multichannel optical spectrometer. The spectrometer system has been equipped with fast time-resolved electronics in order to study the temporal response of the elemental emission lines. The aim of the research was to investigate possible gains in signal-to-background ratio (S/B) and detection limits compared to a conventional direct current (dc) mode of operation. It was found that the high instantaneous pulse power in itself gives approximately a factor 2 increase in S/B. By utilising time-resolved detection, gains in the range 5-10 are achieved. For nitrogen, it was found that the background from leakage of air could be reduced to approximately 1/3 compared with the dc mode. Another interesting observation was that certain nearly non-conducting materials, which cannot be run in dc mode, could be run in the microsecond pulsed mode.