期刊
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES
卷 24, 期 14, 页码 2133-2143出版社
MARCEL DEKKER INC
DOI: 10.1081/JLC-100104897
关键词
-
Collection capacity of lipids from a model fat in supercritical fluid extraction using a solid phase traps was investigated. It was found that 0.6 g of an octadecylsilica material efficiently trapped 80-100 mg of fat. By developing a fractionated extraction/elution procedure, samples containing up to 500 mg of fat could easily be trapped. Losses due to the vapor pressure of the fat components were negligible, even after several hours of purging with gaseous carbon dioxide at a flow rate of ca 1 L/min. The trapping efficiency was found to be independent of the flow rate up to at least ca I L/min of gaseous carbon dioxide, and of modifiers (ethanol, methanol) in concentrations up to at least 10% if the temperature in the trap was sufficiently high to prevent modifier condensation.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐