期刊
ANALYTICAL LETTERS
卷 34, 期 2, 页码 219-229出版社
TAYLOR & FRANCIS INC
DOI: 10.1081/AL-100001574
关键词
trace elements; preconcentration; piperidine dithiocarbamate; inductively coupled plasma atomic emission spectrometry; water
A procedure is developed for the determination of Cd, Co, Cr, Cu, Mn, Ni, Pb, and Zn in water by inductively coupled plasmaatomic emission spectrometry (ICP-AES) after preconcentration on a piperidine dithiocarbamate supported polyurethane foam. The sorbed elements are subsequently eluted with ii mol L-1 HNO3, and the acid eluates are analysed by ICP-AES. A 20-mL disposable syringe served as a chromatographic preconcentration column. The sorption recoveries of elements except Co and Cr were higher than 97%. The method is also applied for the analysis of natural water samples.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐