Atomization of lead from whole blood using novel tungsten filaments in electrothermal atomic absorption spectrometry

期刊

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/b006644n

关键词

-

向作者/读者索取更多资源

Several novel W-filament designs are investigated and compared as atomization devices for blood Pb measurements by atomic absorption spectrometry. Two designs (short and long wire filaments) an based on rigidly-wound W-filaments from Osram, while a third (etched filament) is based on lithographic imaging and photoetching technology. The filaments are positioned in an aluminum atomization cell that is mounted in the optical path of a compact AAS instrument. This instrument is designed specifically for blood lead testing using W-filament technology with a Pb hollow cathode lamp and incorporates a self-reversal background correction system. Whole blood is diluted 1+4 with a phosphate modifier, Triton X-100 and nitric acid, and 15 muL are deposited on the filament using a micropipette. A simple heating program dries, pyrolyzes and atomizes the sample. Integrated absorbance measurements are used for improved precision and detection limits. Matrix-matched Pb standards are necessary for blood lead calibration. Of the three filaments studied, the long wire filament proved most suitable, with a characteristic mass (m(0)) of 200 pg, a method detection limit (3 sigma) of 1-2 mug dL(-1) (05-0.10 mu mol L-1), and a lifetime of the order of 60-70 firings. Method accuracy is verified using blood lead reference materials from the National Institute of Standards and Technology (SRM 955a,b), the Centers for Disease Control (BLLRS) and the New York State Department of Hearth (RMs 028, 040, and 043). For the long wire filament, blood lead results-below 40 mug dL(-1) were mostly within +/- 1 mug dL(-1) of certified values, and within +/- 10% above 40 mug dL(-1); within-run precision was almost always better than +/- 10%. Additional validation is reported using proficiency test materials and human blood specimens. All blood lead results were within the acceptable limits established by regulatory authorities in the US.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据