期刊
BIOSENSORS & BIOELECTRONICS
卷 17, 期 1-2, 页码 53-59出版社
ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/S0956-5663(01)00262-7
关键词
p(HEMA); polypyrrole; amperometric biosensor; glucose oxidase; cholesterol oxidase; galactose oxidase
The present study reports on the use of p(2-hydroxyethyl methacrylate) (pHEMA) in which polypyrrole and various oxidoreductase enzymes were physically entrapped to function as a viable matrix for the construction of clinically important amperometric biosensors. Glucose oxidase. cholesterol oxidase and galactose oxidase biosensors were constructed. Electrode-supported hydrogel films were prepared by UV polymerization of the HEMA component (containing the dissolved enzyme) followed immediately by electrochemical polymerization (+ 0.7V vs. Ag/AgCl) of the pyrrole component within the interstitial spaces of the pre-formed hydrogel network. The optimized glucose oxidase biosensor displayed a wide linear glucose response range (5.0 x 10(-5) to 2.0 x 10(-2) M), a detection limit (3S(y/x)/sensitivity) of 25 muM and a response time of 35-40 s. The analytical recovery of glucose in serum samples ranged from 98 to 102% with mean coefficients of variation of 4.4% (within-day analyses) and 5.1% (day-to-day analyses). All three sensors displayed good stabilities when stored desiccated in the absence of buffer (> 9 months). (C) 2002 Elsevier Science B.V. All rights reserved.