4.4 Article

Comparison of near-infrared transmittance and reflectance measurements using dispersive and Fourier transform spectrophotometers

期刊

METROLOGIA
卷 39, 期 2, 页码 157-164

出版社

BUREAU INT POIDS MESURES
DOI: 10.1088/0026-1394/39/2/5

关键词

-

向作者/读者索取更多资源

A spectrophotometer based on an integrating sphere coupled to a commercial Fourier Transform Infrared (FTIR) instrument has been constructed for regular spectral transmittance and reflectance measurements over the 1 mum to 18 mum wavelength region. Despite the large number of sources of error that often limit the radiometric accuracy of FTIR measurements, we demonstrate uncertainties similar to established dispersive instrumentation. We performed near-normal-incidence reflectance and transmittance measurements on a series of samples over the 1 mum to 2.5 mum wavelength range using both FTIR and dispersive spectrophotometers. The results are compared, taking into account any differences in measurement geometry among the various systems and the combined measurement uncertainty. We find agreement within the combined uncertainties over most of the measured spectral region.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据