4.6 Article

The x-ray sensitivity of amorphous selenium for mammography

期刊

MEDICAL PHYSICS
卷 29, 期 3, 页码 319-324

出版社

WILEY
DOI: 10.1118/1.1449874

关键词

amorphous selenium; x-ray sensitivity; mammography

向作者/读者索取更多资源

A study of the x-ray sensitivity of amorphous selenium (a-Se) for digital mammography has been performed. A uniform layer of a-Se was deposited on a glass substrate with electrodes on both surfaces. The deposition procedure was identical to that used for a-Se flat-panel detectors. A high voltage was applied to the top surface of the a-Se layer in order to establish an electric field E-Se. Then the sample was exposed to x rays with 27 kVp spectra generated from an x-ray tube with a molybdenum (Mo) target. The mean x-ray energy of the spectrum used was similar to16.6 keV. The x-ray current generated by the a-Se layer was measured as a function of E-Se. From the current measurement and the estimation of total x-ray energy absorbed in the a-Se, the energy required to create one electron-hole pair (EHP), W, was determined as a function of E-Se. It was found that at the most commonly used E-Se of 10 V/mum, W was measured as 64 eV. This is considerably higher than the widely accepted typical value of W = 50 eV measured at higher x-ray photon energies (e.g., 50 keV). The dependence of W as a function of E-Se can be best fitted using the empirical expression of E-Se(-y). This relationship is consistent with the results obtained at higher x-ray energies. This article provides an accurate measurement of x-ray sensitivity of a-Se at mammographic energies independent of detector operation, such as the most recently developed flat-panel detectors. The results will be a useful toot for investigation and optimization of a-Se-based x-ray imaging detectors, such as determination of pixel fill-factor and optimal E-Se during operation. (C) 2002 American Association of Physicists in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据