4.8 Article

Demonstration of four immunoassay formats using the array biosensor

期刊

ANALYTICAL CHEMISTRY
卷 74, 期 5, 页码 1061-1068

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac0157268

关键词

-

向作者/读者索取更多资源

The ability of a fluorescence-based array biosensor to measure and quantify the binding of an antigen to an immobilized antibody has been demonstrated using the four different immunoassay formats: direct, competitive, displacement, and sandwich. A patterned array of antibodies specific for 2,4,6-trinitrotoluene (TNT) was immobilized onto the surface of a planar waveguide and used to measure signals from different antigen concentrations simultaneously. For direct, competitive, and displacement assays, which are one-step assays, measurements were obtained in real time. Dose-response curves were calculated for all four assay formats, demonstrating the array biosensor's ability to quantify the amount of antigen present in solution.