4.7 Article

Spiking of performance reference compounds in low density polyethylene and silicone passive water samplers

期刊

CHEMOSPHERE
卷 46, 期 8, 页码 1157-1161

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/S0045-6535(01)00200-4

关键词

water; organic contaminants; SPMD; PRC; passive sampling

向作者/读者索取更多资源

A method for spiking performance reference compounds (PRCs) into low density polyethylene (LDPE) and silicone polymers is described. This method is based on equilibration of the polymers in aqueous/methanolic solutions of PRCs. Equilibration times range from minutes to hours for perdeuterated polyaromatic hydrocarbons (PAHs) and polychlorinated biphenyls (PCBs) with a hydrophobicity in the range 3.9 < logK(ow) < 7.7. Best results were obtained for methanol-water ratios of 80/20 (v/v). The method allows for the use of PRCs for in situ calibration of the uptake kinetics of these solvent-free sampling phases in a similar way as for semipermeable membrane devices (SPMDs). (C) 2002 Elsevier Science Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据