期刊
ANALYTICAL AND BIOANALYTICAL CHEMISTRY
卷 372, 期 2, 页码 284-292出版社
SPRINGER-VERLAG BERLIN
DOI: 10.1007/s00216-001-1128-3
关键词
amperometry; creatinine; creatine; hemodialysis; screen-printed electrode; biosensor
Biosensors for the determination of creatinine have been developed and integrated into a flow-through system. The sensors are based on a screen-printed three electrode transducer with a platinum working electrode. Applying the multi-enzyme sequence of creatininase (CA), creatinase (CI) and sarcosine oxidase (SO) hydrogen peroxide has been detected amperometrically. An optimal enzyme load was found to be 4.4 U/0.28 U/0.20 U (CA/CI/SO) and 0.28 U/0.20 U (CI/SO) per electrode for the creatinine sensor and for the creatine sensor, respectively. Among a variety of polymers Nafion has shown the highest efficiency to exclude interfering substances like ascorbic acid, acetaminophen and uric acid. First determinations of creatinine in dialysate samples obtained during hemodialysis treatments have shown a good correlation to the conventional methods, the Jaffe reaction (y=0.945x+ 2.8, R=0.9882, n=9) and the enzymatic photometric method (y=0.891x+3.5, R=0.9917, n=9).
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据