4.7 Article

Evaluation of an ICP-collision/reaction cell-MS system for the sensitive determination of spectrally interfered and non-interfered elements using the same gas conditions

期刊

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
卷 17, 期 10, 页码 1240-1247

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/b204786c

关键词

-

向作者/读者索取更多资源

Removal of argon-based polyatomics (ArAr and ArO) that interfere with the determination of the main isotopes of selenium and iron was studied using an ICP-MS equipped with a hexapole collision cell. Various gases, including helium, hydrogen, ammonia and N2O, were evaluated to improve detection limits. The behavior of the non-interfered elements was also evaluated using the optimized gas conditions in order to keep the multi-element capability of the ICP-MS. Low detection limits were obtained when a mixture of helium and hydrogen was employed, the values being 2 ng L-1 and 6 ng L-1 for selenium (Se-80) and iron (Fe-56), respectively. The selected conditions did not degrade the LODs of the non-interfered elements.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据