4.6 Article

Selective amperometric detection of dopamine using OPPy-modified diamond microsensor system

期刊

ANALYST
卷 127, 期 12, 页码 1572-1575

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/b208729d

关键词

-

向作者/读者索取更多资源

Highly boron-doped diamond microfiber electrodes (BDDMF) were fabricated and characterized by the use of Scanning Electron Microscopy (SEM), Raman spectroscopy, and cyclic voltammetry. Amperometric detection of dopamine (DA), a neurotransmitter was achieved at pH 7.0, using BDDMF electrodes. The interferences from ascorbic acid ( AA) and DOPAC were efficiently eliminated by using overoxidized polypyrrole-modified BDDMF electrodes, which also increased the sensitivity for the detection of dopamine. The limit of detection (S/N = 3) for dopamine was 0.1 nM, which is one order lower than that observed for carbon microfiber electrodes (CMFE), and the linear dynamic range was obtained from 0.5 nM to 100 muM (r(2) = 0.997). The amperometric response for 0.5 nM dopamine has shown high stability with an RSD of 5.4% ( n = 5). Highly reproducible results were obtained with an RSD of 6.2% for 10 measurements of 1 nM DA taken during 10 h and also remained the same, during measurements for 7 days, with no variation in efficiency for rejection of AA and DOPAC.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据