4.4 Article

Separation/preconcentration of Cu(II), Fe(III), Pb(II), Co(II), and Cr(III) in aqueous samples on cellulose nitrate membrane filter and their determination by atomic absorption spectrometry

期刊

ANALYTICAL LETTERS
卷 35, 期 9, 页码 1561-1574

出版社

TAYLOR & FRANCIS INC
DOI: 10.1081/AL-120006731

关键词

preconcentration; separation; cellulose nitrate membrane; natural water samples; FAAS

向作者/读者索取更多资源

A simple enrichment method for trace amounts of Cu(II), Fe(III), Pb(II), Co(II) and Cr(III) has been established. The preconcentration/separation procedure is based on chelate formation of the analyte metals with benzopurpurine 4B and on retention of the chelates on cellulose nitrate membrane filter. After completely dissolving the cellulose nitrate membrane and analyte ions by 500 muL of nitric acid at 85degreesC on a hood, metal determinations were performed by flame atomic absorption spectrometry. The influences of some analytical parameters including pH, types of membrane filters, amounts of reagent etc. on the recoveries of copper, iron, lead, cobalt and chromium were investigated. The detection limits of the analyte ions (k = 3, N = 2 1) were varying 1.9 mug/L for cobalt to 10.5 mug/L for iron. The method was applied to natural water samples for the determination of analyte ions with satisfactory results (recoveries > 95%, RSD < 10%). The relative standard deviations (RSD) of the determinations for analyte ions were below 10%.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据