4.4 Article

Validity and reliability of three commercially available breath-by-breath respiratory systems

期刊

EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY
卷 86, 期 5, 页码 435-441

出版社

SPRINGER-VERLAG
DOI: 10.1007/s00421-001-0572-2

关键词

gas exchange; exercise testing; reproducibility; breath-by-breath measurements; on-line systems

向作者/读者索取更多资源

Information concerning the validity and reliability of commercial on-line gas analysis systems is limited. The aim of this study was to provide a comparison of the validity and reliability of three on-line systems (Oxycon Alpha, Oxycon Pro and Pulmolab EX670) with that of Douglas bags. Two separate studies were conducted. In study 1, the three gas analysis systems were compared with Douglas bags using a metabolic simulator over four increases in ventilation. In study 2, ten subjects were split into equal groups exercising at 100 W or 150 W for 85 min on three separate occasions. Each system was used twice per visit. Study I demonstrated that the Oxycon Alpha and Douglas bags produced similar respiratory values over all levels of ventilation. The Oxycon Pro tended to slightly overestimate mean expiratory flow ((V) over dot (E)), oxygen uptake ((V) over dot O-2), carbon dioxide production ((V) over dot CO2) and respiratory exchange ratio (RER) at the higher ventilations The Pulmolab produced large overestimations at all ventilations for (V) over dot CO2 and RER (up to. 26.3% away from expectations), whilst values for (V) over dot (E), and (V) over dot O-2 were slightly underestimated at higher ventilations (up to 7.5% from expectations). The results of study 2 support the findings of study 1, with the Oxycon Pro and Oxycon Alpha producing. similar results compared to Douglas bags for (V) over dot O-2, (V) over dot CO2 and RER. The coefficients of variation for (V) over dot O-2 and (V) over dot CO2 measured using Douglas bags, Oxycon Pro and Oxycon Alpha were 3.3-5.1%, 4.7-7.0% and 4.5-6.3%, respectively, whilst that for the Pulmolab was highly variable (26.8-45.8%). The exercise study showed the Oxycon Pro and Oxycon Alpha to be both valid and reliable on-line systems for the measurement of parameters of respiration, at least at workloads up to 150 W.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据