4.6 Article

First clinical use of a novel bioartificial liver support system (BLSS)

期刊

AMERICAN JOURNAL OF TRANSPLANTATION
卷 2, 期 3, 页码 260-266

出版社

BLACKWELL MUNKSGAARD
DOI: 10.1034/j.1600-6143.2002.20311.x

关键词

bioartificial liver; liver failure; porcine hepatocytes

向作者/读者索取更多资源

The first clinical use of the Excorp Medical Bioartificial Liver Support System (BLSS) in support of a 41-year-old African-American female with fulminant hepatic failure is described. The BLSS is currently in a Phase I/II safety evaluation at the University of Pittsburgh/UPMC System. Inclusion criteria for the study are patients with acute liver failure, any etiology, presenting with encephalopathy deteriorating beyond Parson's Grade 2. The BLSS consists of a blood pump; a heat exchanger to control blood temperature; an oxygenator to control oxygenation and pH; a bioreactor; and associated pressure and flow alarm systems. Patient liver support is provided by 70-100 g of porcine liver cells housed in the hollow fiber bioreactor. The patient exhibited transient hypotension and thrombocytopenia at initiation of perfusion. The only unanticipated safety event was a lowering of patient glucose level at the onset of perfusion with the BLSS that was treatable with intravenous glucose administration. Moderate changes in blood biochemistries pre- and post perfusion are indicative of liver support being provided by the BLSS. While the initial experience with the BLSS is encouraging, completion of the Phase I/II study is required in order to more fully understand the safety aspects of the BLSS.