4.5 Article

The effect of cleaning procedures on fracture properties and corrosion of NiTi files

期刊

INTERNATIONAL ENDODONTIC JOURNAL
卷 36, 期 11, 页码 724-732

出版社

BLACKWELL PUBLISHING LTD
DOI: 10.1046/j.1365-2591.2003.00709.x

关键词

cleaning procedures; corrosion; fracture properties; NiTi files

向作者/读者索取更多资源

Aim To evaluate the effect of repeated cleaning procedures on fracture properties and corrosion of nickel-titanium (NiTi) files. Methodology New NiTi instruments were subjected to 2, 5 and 10 cleaning cycles with the use of either diluted bleach (1% NaOCl) or Milton's solution (1% NaOCl plus 19% NaCl) as disinfectant. Each cleaning cycle consisted of scrubbing, rinsing and immersing in NaOCl for 10 min followed by 5 min of ultrasonication. Files were then tested for torsional failure and flexural fatigue, and observed for evidence of corrosion using scanning electron microscope (SEM). Four brands of NiTi files were immersed in either Milton's solution or diluted bleach overnight and evaluated for corrosion. Results Up to 10 cleaning cycles did not significantly reduce the torque at fracture or number of revolutions to flexural fatigue (P > 0.05, two-way ANOVA), although decreasing values were noted with increasing number of cleaning cycles using Milton's solution. No corrosion was detected on the surface of these files. Files immersed in 1% NaOCl overnight displayed a variety of corrosion patterns. The extent of corrosion was variable amongst different brands of files and amongst files in each brand. Overall, Milton's solution was much more corrosive than diluted bleach. Corrosion of file handles was often extreme. Conclusions Files can be cleaned up to 10 times without affecting fracture susceptibility or corrosion, but should not be immersed in NaOCl overnight. Milton's solution is much more corrosive than bleach with the same NaOCl concentration.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据