4.7 Article

Isocitrate analysis using a potentiometric biosensor with immobilized enzyme in a FIA system

期刊

FOOD RESEARCH INTERNATIONAL
卷 36, 期 3, 页码 223-230

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0963-9969(02)00140-0

关键词

isocitrate; biosensor; enzyme immobilization; flow injection analysis (FIA)

向作者/读者索取更多资源

A potentiometric biosensor for the analysis of isocitrate was developed by using a CO32--selective electrode and enzyme immobilization in flow injection analysis (FIA). The biosensor consisted of a peristaltic pump, injector, enzyme reactor, reference electrode, working electrode, pH/mV meter and record. The experimental parameters that influenced the sensitivity and selectivity of the biosensors were optimized in the FIA system. A linear correlation between the potential difference and logarithmic isocitrate concentration was obtained in the range of 10(-3)-10(-1) M isocitrate under optimal conditions. The interference effect of major sugars and organic acids on the sensor system was less than 5%. Isocitrate concentrations of some fruit juices analyzed by the isocitrate sensor system were compared with those analyzed by gas chromatography (GC). There was no significant difference between the two analytical methods in any of the fruit juices. This suggests that the isocitrate sensor system is reliable in determining the isocitrate concentrations of foods. (C) 2002 Elsevier Science Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据