4.4 Article

Calibration of the Orsted vector magnetometer

期刊

EARTH PLANETS AND SPACE
卷 55, 期 1, 页码 11-18

出版社

SPRINGEROPEN
DOI: 10.1186/BF03352458

关键词

-

向作者/读者索取更多资源

The vector fluxgate magnetometer of the Orsted satellite is routinely calibrated by comparing its output with measurements of the absolute magnetic intensity from the Overhauser instrument, which is the second magnetometer of the satellite. We describe the method used for and the result obtained in that calibration. Using three years of data the agreement between the two magnetometers after calibration is 0.33 nT rms (corresponding to better than +/-1 nT for 98% of the data, and better than +/-2 nT for 99.94% of the data). We also report on the determination of the transformation between the magnetometer coordinate system and the reference system of the star imager. This is done by comparing the magnetic and attitude measurements with a model of Earth's magnetic field. The Euler angles describing this rotation are determined in this way with an accuracy of better than 4 arcsec.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据