The simultaneous use of a biventricular implantable cardioverter defibrillator (ICD) and transcutaneous electrical nerve stimulation (TENS) unit: implications for device interaction

期刊

EUROPACE
卷 5, 期 1, 页码 91-93

出版社

W B SAUNDERS CO LTD
DOI: 10.1053/eupc.2002.0277

关键词

transcutaneous electrical nerve stimulation; implantable cardioverter defibrillator; device interaction

向作者/读者索取更多资源

Undesirable sensing of external sources of electromagnetic interference by ICDs is well known. A transcutaneous electrical nerve stimulation (TENS) device has been reported to interfere with an ICD resulting in an inappropriate shock and patients with implanted defibrillators or pacemakers have been cautioned about the use of such units to treat chronic pain. We describe a patient regularly using TENS therapy for pain who subsequently received a biventricular ICD for malignant ventricular arrhythmias and medically refractory cardiac failure. He underwent testing for device interaction immediately post-implant. This did not show inappropriate sensing by either ICD or pacemaker component of his heart failure device. However, six months later, the patient complained of dizziness and bradycardia with application of TENS. Further testing did reveal interference with pacemaker function. Thus, even if initial testing is negative and reassuring, patients with a biventricular ICD still require careful follow-up for potential interaction and should be cautioned against the use of TENS, especially if they are pacemaker-dependent. (C) 2003 The European Society of Cardiology. Published by Elsevier Science Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据