4.3 Article

Adapting the Quesant (c) Nomad (TM) atomic force microscope for biology and patch-clamp atomic force microscopy

期刊

CELL BIOCHEMISTRY AND BIOPHYSICS
卷 39, 期 3, 页码 195-210

出版社

HUMANA PRESS INC
DOI: 10.1385/CBB:39:3:195

关键词

AFM; phase contrast; patch-clamp AFM; electrophysiology; electro notility

向作者/读者索取更多资源

The Quesant(C) Nomad(TM) atomic force microscope (AFM) was modified to produce a reliable patch-clamp AFM for demanding biologic applications. The AFM's laser optics forms the basis of a condenser that allows simultaneous Kohler illumination and AFM imaging on an inverted optical microscope. The original AFM scan head was replaced with plastic and glass to make it biologically inert. A bevel cut in the new scan head permits clearance for patch clamp pipets. Cantilevers are attached to the scan head with a quick setting silicone rubber that is readily removable. Software was developed to (a) automate a gentle approach and set a specific feedback force, (b) provide a mouse-driven control of the X-Y position of the probe tip and recall of saved locations, and (c) measure force-distance curves over user defined paths. Additional modifications were made to minimize mechanical noise. The patch-clamp AFM achieves 600 fA (3 kHz bandwidth) and 1 Angstrom RMS noise levels (10 kHz bandwidth). The correlation of electrical and mechanical information allows signal averaging and measures sub-Angstrom, sub-millisecond electromotile responses from cells.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据