期刊
出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.nima.2003.08.059
关键词
pixel array detector; X-ray absorption spectroscopy; fluorescence detection; synchrotron radiation
For high throughput/energy-resolution fluorescence X-ray detection in X-ray absorption spectroscopy (XAS), a novel Ge pixel array detector (PAD) with 100 segments has been developed. Using a monolithic approach, a high packing density (88%) with nearly perfect commission rate (99%) were achieved, resulting in a dramatic improvement in packing density compared to the close-packed multi-element detector (57%). Two types of PAD was built with the pixel thickness of 7.5 and 10 mm. Each segment has an active area of 22 mm(2). The average energy resolution at 5.9 keV was ca. 220 eV (6 mus shaping time) and 240 eV (0.5 mus shaping time). High efficiency in X-ray detection in a wide energy range (5-60 keV) was confirmed. The present data acquisition system, CAMAC-based hybrid electronics, is capable of independent energy-analysis and recording all channels with a maximum 100 kcps per channel or 10 MHz in total. Elimination of elastic and inelastic scattering improved the systematic error in XAS, as demonstrated by an XAS application to a photo-induced phase transition. (C) 2003 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐