Radiation damage and charge collection effects in S(Li) gamma-ray detectors

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0168-9002(02)01619-4

关键词

Si(Li) detector; radiation damage; gamma ray

向作者/读者索取更多资源

The spectroscopy performance of 6-mm thick Si(Li) planar detectors was studied as a function of operating temperature and electric field. The energy resolution of the 662-keV gamma-ray peak from a Cs-137 source was used to monitor the spectroscopy performance of the detectors. The efficiency, depletion voltage, leakage current, and noise were also monitored. The effects of radiation damage caused by 200-MeV protons were studied to determine the viability of operation in space. Four detectors, two maintained at 88 K and two maintained at 212 K, were irradiated to a fluence of 8.7 x 10(8)p/cm(2). No effects were observed. The two detectors irradiated at 212K were subsequently irradiated with an additional 8.7 x 10(9) p/cm(2), again at 212 K. These detectors then exhibited slight energy resolution degradation. No other radiation damage effects were observed. The resolution degradation increased at higher operating temperature and decreased with higher electric field. Cycling the detectors to room temperature for 14 h eliminated the resolution degradation. The resolution of these detectors is limited by the combination of ballistic deficit and parallel noise in the similar to220 K range. A significant decrease in the gamma-ray peak count rate, almost certainly caused by surface channel effects, was the most dramatic temperature-dependent effect observed. (C) 2002 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据