Aneurysm of the right ventricular outflow following bovine valved venous conduit insertion

期刊

EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY
卷 23, 期 1, 页码 122-124

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S1010-7940(02)00660-7

关键词

pulmonary valved conduit; Contegra (R) bovine jugular vein graft; homograft; xenograft; right ventricular outflow tract reconstruction; aneurysm

向作者/读者索取更多资源

A case of aneurysm of the right ventricular outflow tract is described after repair of tetralogy of Fallot using a Contegra supported conduit. Angiograms revealed that the aneurysm was located between the ventricular anastomosis and the proximal ring of the conduit confirming echocardiographic data. Because the conduit between the rings was not dilated, the valve was perfectly functioning. Pulmonary anastomosis was severely stenosed explaining the dilatation seen below. Conduit replacement with resection of the aneurysmal part of the failing conduit was performed. Supported conduits do not eliminate the risk of secondary dilatation below the artificial ring but preserve valvular function. (C) 2002 Elsevier Science B.V. All rights reserved.