RAPID COMMUNICATIONS IN MASS SPECTROMETRY
卷 17, 期 5, 页码 429-436出版社
JOHN WILEY & SONS LTD
DOI: 10.1002/rcm.927
关键词
-
We propose a new algorithm for deconvolution of electrospray ionization mass spectra based on direct assignment of charge to the measured signal at each mass-to-charge ratio (m/z). We investigate two heuristics for charge assignment: the entropy-based heuristic is adapted from a deconvolution algorithm by Reinhold and Reinhold;(10) the multiplicative-correlation heuristic is adapted from the multiplicative-correlation deconvolution algorithm of Hagen and Monnig.(6) The entropy-based heuristic is insensitive to overestimates of z(max), the maximum ion charge. We test the deconvolution algorithm on two single-component samples: the measured spectrum of human beta-endorphin has two prominent and one very weak line whereas myoglobin has a well-developed quasi-gaussian envelope of 17 peaks. In both cases, the deconvolution algorithm gives a clean deconvoluted spectrum with one dominant peak and very few artefacts. The relative heights of the peaks due to the parent molecules in the deconvoluted spectrum of a mixture of two peptides, which are expected to ionize with equal efficiency, give an accurate measure of their relative concentration in the sample. Copyright (C) 2003 John Wiley Sons, Ltd.