4.4 Article

Heuristic charge assignment for deconvolution of electrospray ionization mass spectra

期刊

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
卷 17, 期 5, 页码 429-436

出版社

JOHN WILEY & SONS LTD
DOI: 10.1002/rcm.927

关键词

-

向作者/读者索取更多资源

We propose a new algorithm for deconvolution of electrospray ionization mass spectra based on direct assignment of charge to the measured signal at each mass-to-charge ratio (m/z). We investigate two heuristics for charge assignment: the entropy-based heuristic is adapted from a deconvolution algorithm by Reinhold and Reinhold;(10) the multiplicative-correlation heuristic is adapted from the multiplicative-correlation deconvolution algorithm of Hagen and Monnig.(6) The entropy-based heuristic is insensitive to overestimates of z(max), the maximum ion charge. We test the deconvolution algorithm on two single-component samples: the measured spectrum of human beta-endorphin has two prominent and one very weak line whereas myoglobin has a well-developed quasi-gaussian envelope of 17 peaks. In both cases, the deconvolution algorithm gives a clean deconvoluted spectrum with one dominant peak and very few artefacts. The relative heights of the peaks due to the parent molecules in the deconvoluted spectrum of a mixture of two peptides, which are expected to ionize with equal efficiency, give an accurate measure of their relative concentration in the sample. Copyright (C) 2003 John Wiley Sons, Ltd.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据