MAGNETIC RESONANCE IN MEDICINE
卷 50, 期 5, 页码 1018-1030出版社
JOHN WILEY & SONS INC
DOI: 10.1002/mrm.10597
关键词
SNR; parallel imaging; RF coil arrays
资金
- NATIONAL HEART, LUNG, AND BLOOD INSTITUTE [R29HL060802] Funding Source: NIH RePORTER
- NATIONAL INSTITUTE OF BIOMEDICAL IMAGING AND BIOENGINEERING [R01EB000447] Funding Source: NIH RePORTER
- NHLBI NIH HHS [R29 HL60802] Funding Source: Medline
- NIBIB NIH HHS [R01 EB00447] Funding Source: Medline
A method is described for establishing an upper bound on the spatial encoding capabilities of coil arrays in parallel MRI. Ultimate intrinsic signal-to-noise ratio (SNR), independent of any particular conductor arrangement, is calculated by expressing arbitrary coil sensitivities in terms of a complete set of basis functions that satisfy Maxwell's equations within the sample and performing parallel imaging reconstructions using these basis functions. The dependence of the ultimate intrinsic SNR on a variety of experimental conditions is explored and a physically intuitive explanation for the observed behavior is provided based on a comparison between the electromagnetic wavelength and the distance between aliasing points. Imaging at high field strength, with correspondingly short wavelength, is shown to offer advantages for parallel imaging beyond those already expected due to the larger available spin polarization. One-dimensional undersampling of k-space yields a steep drop in attainable SNR for more than a 5-fold reduction of scan time, while 2D undersampling permits access to much higher degrees of acceleration. Increased tissue conductivity decreases baseline SNR, but improves parallel imaging performance. A procedure is also provided for generating the optimal coil sensitivity pattern for a given acceleration, which will serve as a useful guide for future coil designs. Magn Reson Med 50: 1018-1030, 2003. (C) 2003 Wiley-Liss, Inc.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据