4.6 Article

Parallel detection of five human herpes virus DNAs by a set of real-time polymerase chain reactions in a single run

期刊

JOURNAL OF CLINICAL VIROLOGY
卷 26, 期 1, 页码 85-93

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S1386-6532(02)00042-2

关键词

cytomegalovirus (CMV); Epstein-Barr virus (EBV); herpes simplex virus (HSV); varicella-zoster virus (VZV); real-time polymerase chain reaction (PCR); LightCycler

类别

向作者/读者索取更多资源

Background: Human herpes viruses cause a spectrum of diseases that are usually self-limiting but can be reactive during immuno-suppression and may then lead to severe or even life-threatening diseases. The LightCycler technology allows rapid polymerase chain reaction (PCR) including product analysis within a closed system. This approach has been demonstrated to be suitable for routine diagnostic virus detection. Several LightCycler PCR assays have been established to the detection of human herpes viruses. The assays vary in their detection formats and PCR cycling protocols. So, they cannot be performed within a single LightCycler run. Objectives: Development of four LightCycler PCR assays for parallel detection of DNA derived from human cytomegalovirus (CMV), Epstein-Barr virus (EBV), herpes simplex virus type 1 (HSV-1), herpes simplex virus type 2 (HSV-2), and varicella-zoster virus (VZV) in a single run. Study design: Primers and hybridization probes were tailored to suit one LightCycler PCR program. LightCycler PCRs were established, detection limits were determined, and clinical samples were evaluated. Results: With quantified herpes virus type specific DNA spiked into cerebrospinal fluid, serum or EDTA plasma the detection limits were found either at 500 or 250 viral DNA copies per ml depending on the virus DNA specific PCR, and on the specimen type used. The applicability of the new LightCycler assays for routine molecular testing was evaluated by testing 96 clinical samples. Conclusion: The developed set of LightCycler PCRs permits parallel detection of CMV, EBV, HSV-1, HSV-2, and VZV in a single LightCycler run. The new molecular assays can easily be used to the rapid, simple, and convenient detection of herpes virus DNA in cerebrospinal fluid, serum and EDTA plasma in the routine diagnostic laboratory. (C) 2002 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据