4.6 Article

Determination of erythrocyte deformability and its correlation to cellular ATP release using microbore tubing with diameters that approximate resistance vessels in vivo

期刊

ANALYST
卷 128, 期 9, 页码 1163-1168

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/b308225n

关键词

-

资金

  1. NATIONAL HEART, LUNG, AND BLOOD INSTITUTE [R01HL073942] Funding Source: NIH RePORTER
  2. NHLBI NIH HHS [HL 073942-01] Funding Source: Medline

向作者/读者索取更多资源

A novel method is described for measuring the deformability of red blood cells (RBCs) in tubing whose diameters approximate forces encountered in vivo. Here, RBCs from rabbits are loaded into a 50 cm section of 75 mm id microbore tubing and connected to a syringe pump. This section of tubing is then connected to a 15 cm section of 25 mum id tubing. As buffer is pumped through the flow system, the RBCs are evacuated from both sections of tubing. However, the inability of the RBCs to move freely through the 25 mum id section of tubing results in a buildup of cells at the inlet of this portion of tubing. The continued force output by the syringe pump results in a deformation of the RBCs until all of the cells are eventually evacuated from the flow system. It was found that a measurement of the time required to reach half of the maximum pressure (1/2P(max)) may be used as an indicator of the RBC deformability. For a given sample, a simple buffer results in less time to reach 1/2P(max) (6.9 +/- 0.2 s) than deformable RBCs (21.6 +/- 0.8 s). To verify that the increased amount of time to reach 1/2P(max) is indeed due to the RBCs, various hematocrits of an RBC sample were investigated and, as expected, it was found that a 12% RBC hematocrit had a higher 1/2P(max) value (26.0 s +/- 2.2 s) when compared to a 7% hematocrit (19.1 +/- 0.3 s). In addition, RBCs chemically stiffened with glutaraldehyde were shown to be 25% less deformable than normal RBCs. Finally, a study was performed to examine the relationship between RBC deformability and ATP release and it was found that ATP release increased as a function of RBC deformability. This method greatly simplifies deformability measurements, employing only a syringe pump and microbore tubing, and may lead to a more complete understanding of the physiological significance of erythrocyte deformability.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据