In vitro evaluation of ultrasound Doppler strain rate imaging: Modification for measurement in a slowly moving tissue phantom

期刊

ULTRASOUND IN MEDICINE AND BIOLOGY
卷 29, 期 12, 页码 1725-1734

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ultrasmedbio.2003.08.006

关键词

strain rate imaging; Doppler ultrasound; in vitro; silicone rubber strip; smooth muscle; gastric wall; gastric antrum; gastrointestinal tract

向作者/读者索取更多资源

Doppler strain rate imaging (SRI) was evaluated in vitro using a silicone strip phantom mimicking slowly moving tissue. A test apparatus was developed that enabled controlled strain experiments with variable strain and strain rate to be performed. SRI strain was measured at eight different calculated strains (range 5.7 to 63.4 %) at three different pump speeds with tissue velocity 0.1, 0.5 and 1.0 mm/s. The effect of varying tissue velocity and strain sample size on the measured SRI strain was elaborated. SRI strains agreed well with calculated values for strain when SRI strain was measured as the average over the whole strip cross-section and the strain sample size was 1.9 mm (mean difference = 2.78%, limits of agreement +/- 9.97% for tissue velocity 1.0 mm/s, n = 8). The variance was substantial if single central samples were used, especially for strain sample size of 0.8 mm (mean difference = -7.47%, limits of agreement +/- 20.90 for tissue velocity 0.5 mm/s, n = 24). Increasing the strain sample size to 1.9 mm removed some of the underestimation (giving mean difference of -4.46%, n = 24). We found low intra- and interobserver variation. This study indicates that, for the SRI method to give accurate estimates of strain, strain sample size should be in the region of 2 mm. Averaging over several ultrasound (US) beams increased the accuracy further.(E-mail: Knut.Matre@med.uib.no) (C) 2003 World Federation for Ultrasound in Medicine Biology.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据