4.5 Article

Practical limits of resolution in confocal and non-linear microscopy

期刊

MICROSCOPY RESEARCH AND TECHNIQUE
卷 63, 期 1, 页码 18-22

出版社

WILEY
DOI: 10.1002/jemt.10423

关键词

confocal; second harmonic; multiphoton; resolution; super-resolution

向作者/读者索取更多资源

Calculated and measured resolution figures are presented for confocal microscopes with different pinhole sizes and for nonlinear (2-photon and second harmonic) microscopes. A modest degree of super-resolution is predicted for a confocal microscope but in practice this is not achievable and confocal fluorescence gives little resolution improvement over widefield. However, practical non-linear microscopes do approach their theoretical resolution and therefore show no resolution disadvantage relative to confocal microscopes in spite of the longer excitation wavelength.