期刊
ANTI-CORROSION METHODS AND MATERIALS
卷 51, 期 5, 页码 331-338出版社
EMERALD GROUP PUBLISHING LTD
DOI: 10.1108/00035590410554892
关键词
corrosion; bacteria; medical appliances
Microbiologically induced corrosion of copper intrauterine devices TCu380A and Nova T by Enterobacter sp. in a synthetic intrauterine medium was investigated. The corrosion behaviour was evaluated applying cyclic voltammetry and Tafel Polarization. The surface morphology was determined using scanning electron microscopy, the corrosion products were analyzed using energy dispersive X-rays and the biofilm composition was examined by Fourier transform infrared spectroscopy with an attenuated total reflectance device.