4.5 Article

Measurement of C-reactive protein: Two high sensitivity methods compared

期刊

JOURNAL OF CLINICAL LABORATORY ANALYSIS
卷 18, 期 5, 页码 280-284

出版社

WILEY
DOI: 10.1002/jcla.20038

关键词

atherosclerosis; C-reactive protein high-sensitivity; immunoassay; nephelometry; turbidimetry

向作者/读者索取更多资源

C-reactive protein (CRIP) is an acute phase marker and a predictor of the risk of developing atherosclerotic complications. However, as a predictor of this risk, high sensitivity measurements are needed, and high sensitive CRP (hsCRP) assays have been developed. In this study, we experimentally compared two hsCRP assays, based on nephelometry and turbidimetry, both implemented on automated analyzers. Linearity, imprecision, turbidity interference, and results in the assay of 96 samples have been compared. Method comparison of the same two analytical systems in the assay of CRP was also performed on the basis of results in an interlaboratory external quality assessment scheme (EQAS). The two systems were found to perform substantially equally, both in hsCRP and in CRP measurement, but in the hsCRP assay the precision of nephelometry (CV% in the interval 3.0-5.8) was lower than that of turbidimetry (CV% in the interval 1.8-2.3). The classification of results by the two methods into three predefined relative risk classes gave 18% rate of discordance, in any case by one class only. The two methods proved reliable and comparable in the measurement of hsCRP, but precision should be improved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据