4.1 Article

Frameless stereotactic navigation in transsphenoidal surgery: Comparison with fluoroscopy

期刊

STEREOTACTIC AND FUNCTIONAL NEUROSURGERY
卷 82, 期 1, 页码 43-48

出版社

KARGER
DOI: 10.1159/000076660

关键词

fluoroscopy; frameless stereotactic localization; intraoperative localization; transsphenoidal surgery; viewscope

向作者/读者索取更多资源

Surgical navigation systems ( frameless stereotaxy) have been used in addition to or instead of fluoroscopy during transsphenoidal surgery. This study compares the intraoperative localization by an optical tracking system (Elekta Viewscope) with fluoroscopy. Viewscope and fluoroscope sagittal images were compared by the establishment of a Cartesian coordinate system based on anatomical landmarks and by the spatial localization of surgically relevant points for 20 patients. The Viewscope was found to have a total deviation of 3.0 +/- 0.6 mm (mean +/- SD) compared to fluoroscopy (p < 0.01). Much of the error resulted from the registration process, which according to the Viewscope software had an expected error of 3.1 +/- 0.8 mm for this series of patients, and from the probe-to-system correlation ( error of 1.0 +/- 0.3 mm). Although frameless stereotactic systems give the surgeon useful trajectory data with three-dimensional visualizations, they remain somewhat inaccurate. The multiplanar abilities of the Viewscope provide an additional but not mandatory advantage to the simplicity and accuracy of fluoroscopy during this type of surgery. Copyright (C) 2004 S. Karger AG, Basel.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据