4.1 Article

Canine and feline hematology reference values for the ADVIA 120 hematology system

期刊

VETERINARY CLINICAL PATHOLOGY
卷 33, 期 1, 页码 32-38

出版社

AMER SOC VETERINARY CLINICAL PATHOLOGY
DOI: 10.1111/j.1939-165X.2004.tb00347.x

关键词

ADVIA 120; blood cell morphology; cat; complete blood count; dog; hematology; reference interval

向作者/读者索取更多资源

Background: The ADVIA 120 is a laser-based hematology analyzer with software applications for animal species. Accurate reference values would be useful for the assessment of new hematologic parameters and for interlaboratory comparisons. Objective: The goal of this study was to establish reference intervals for CBC results and new parameters for RBC morphology, reticulocytes, and platelets in healthy dogs and cats using the ADVIA 120 hematology system. Methods: The ADVIA 120, with multispecies software (version 1.107-MS), was used to analyze whole blood samples from clinically healthy dogs (n=46) and cats (n=61). Data distribution was determined and reference intervals were calculated as 2.5 to 97.5 percentiles and 25 to 75 percentiles. Results: Most data showed Gaussian or log-normal distribution. The numbers of RBCs falling outside the normocyticnormochromic range were slightly higher in cats than in dogs. Both dogs and cats had reticulocytes with low, medium, and high absorbance. Mean numbers of large platelets and platelet clumps were higher in cats compared with dogs. Conclusions: Reference intervals obtained on the ADVIA 120 provide valuable baseline information for assessing new hematologic parameters and for interlaboratory comparisons. Differences compared with previously published reference values can be attributed largely to differences in methodology. (C) 2004 American Society for Veterinary Clinical Pathology

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据