4.5 Article

Bolometric technique for high-resolution broadband microwave spectroscopy of ultra-low-loss samples

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 75, 期 1, 页码 124-135

出版社

AMER INST PHYSICS
DOI: 10.1063/1.1633001

关键词

-

向作者/读者索取更多资源

A novel low-temperature bolometric method has been devised and implemented for high-precision measurements of the microwave surface resistance of small single-crystal platelet samples having very low absorption, as a continuous function of frequency. The key to the success of this nonresonant method is the in situ use of a normal metal reference sample that calibrates the absolute rf field strength. The sample temperature can be controlled independently of the 1.2 K liquid-helium bath, allowing for measurements of the temperature evolution of the absorption. However, the sensitivity of the instrument decreases at higher temperatures, placing a limit on the useful temperature range. Using this method, the minimum detectable power at 1.3 K is 1.5 pW, corresponding to a surface resistance sensitivity of approximate to1 muOmega for a typical 1 mmx1 mm platelet sample. (C) 2004 American Institute of Physics.