4.6 Article

Multicenter evaluation of a new immunoassay for procalcitonin measurement on the Kryptor (R) System

期刊

CLINICAL CHEMISTRY AND LABORATORY MEDICINE
卷 42, 期 4, 页码 440-449

出版社

WALTER DE GRUYTER & CO
DOI: 10.1515/CCLM.2004.077

关键词

automation; homogenous immunoassay; procalcitonin

向作者/读者索取更多资源

We compared the manually performed LUMItest(R) procalcitonin (PCT) assay with the newly developed fully mechanized Kryptor(R) PCT assay and determined the essential assay characteristics of this assay. The new Kryptor(R) PCT assay was evaluated according to modified NCCLS EP-10/EP-6 protocols in five different laboratories. Samples from 696 patients were assayed using the original LUMItest(R) PCT assay and the new Kryptor(R) PCT assay. Possible interference by hemoglobin, triglycerides and bilirubin was evaluated by spiking patient plasma with the appropriate substances. The functional assay sensitivity (FAS) was determined by analyzing samples with low PCT concentrations. The FAS of the new Kryptor(R) PCT assay was 0.04 ng/ml and the imprecision within and betweenseries below 5% and below 10%, respectively. Within the smallest range of determination, from 0.3 ng/ml to 50 ng/ml, common to the LUMItest(R) PCT assay (x) and the Kryptor(R) PCT assay (y) the values correlated well: y=0.64+0.94x, s.(xy)=2.78 ng/ml. The performance characteristics of the Kryptor(R) PCT assay are fully compatible with the intended clinical use. The assay allows determination of PCT in a turnaround time (TAT) of about 20 minutes and thus is adequate for STAT analyses.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据