4.3 Article

A miniaturized FIA system for the determination of residual chlorine in environmental water samples

期刊

ANALYTICAL SCIENCES
卷 20, 期 1, 页码 205-207

出版社

JAPAN SOC ANALYTICAL CHEMISTRY
DOI: 10.2116/analsci.20.205

关键词

-

向作者/读者索取更多资源

The present study was aimed at developing a portable miniaturized FIA system with a two-electrode based amperometric detector for on-site monitoring of residual chlorine in environmental samples. The electrode potential control and current data collection were conducted with a home-designed potentiostat with a size of 150 x 100 x 40 mm and the weight of 350 g. It was confirmed that the amperometric response increased in proportion to the chlorine concentration up to 5 mug/ml, and the detection limit for chlorine was 0.05 mug/ml. Besides a drastic reduction in the size of the analytical system, such miniaturization achieved the advantages of increased speed, minimal mobile-phase consumption, and high sensitivity. The proposed method was successfully applied to the determination of residual chlorine in samples of pool water and tap water. The results were well agreed with those obtained by the N,N-diethyl-p-phenylenediamine (DPD) photometric method.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据