3.9 Article

Treatment of unstable trochanteric fractures - Randomised comparison of the gamma nail and the proximal femoral nail

期刊

出版社

BRITISH EDITORIAL SOC BONE & JOINT SURGERY
DOI: 10.1302/0301-620X.86B1.14455

关键词

-

向作者/读者索取更多资源

The proximal femoral nail (PFN) is a recently introduced intramedullary system, designed to improve treatment of unstable trochanteric fractures of the hip. In a multicentre prospective clinical study, the intra-operative use, complications and outcome of treatment using the PFN (n = 211) were compared with those using the gamma nail (GN) (n = 213). The intra-operative blood loss was lower with the PFN (220 ml v 287 ml, p = 0.001). Post-operatively, more lateral protrusion of the hip screws of the PFN (7.6%) was documented, compared with the gamma nail (1.6%, p = 0.02). Most local complications were related to suboptimal reduction of the fracture and/or positioning of the implant. Functional outcome and consolidation were equal for both implants. Generally, the results of treatment of unstable trochanteric fractures were comparable for the PFN and GN. The pitfalls and complications were similar, and mainly surgeon- or fracture-related, rather than implant-related.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.9
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据