期刊
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
卷 19, 期 8, 页码 945-949出版社
ROYAL SOC CHEMISTRY
DOI: 10.1039/b406333c
关键词
-
The new microplasma detector consists of a simple arrangement of two cylindrical electrodes placed on a fused silica capillary, which carry an ac-voltage of 20 kHz and 20 kV to sustain the discharge, and an optical fibre for collection of the emitted light. The detection of Br, Cl, F, I, P, Se and S in organic compounds separated in a helium carrier was possible by monitoring emission lines at 827.24 nm, 837.59 nm, 685.60 nm, 804.37 nm, 956.40 nm, 888.50 nm and 921.28 nm for the 7 elements, respectively. The system showed the following detection limits: Br 0.3, Cl 0.1, F 20, I 158, P 1805, Se 153 and S 6.6 pg s(-1). The determination of environmentally relevant halogenated volatile organic compounds and pesticides is demonstrated.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐