4.4 Article

Solid phase extraction of Cu(II), Pb(H), Fe(III), Co(II), and Cr(III) on Chelex-100 column prior to their flame atomic absorption spectrometric determinations

期刊

ANALYTICAL LETTERS
卷 37, 期 6, 页码 1203-1217

出版社

MARCEL DEKKER INC
DOI: 10.1081/AL-120034064

关键词

Chelex-100; column; solid phase extraction; microwave digestion; wet digestion; AAS

向作者/读者索取更多资源

A solid phase extraction procedure for flame atomic absorption spectrometric determination of Cu(II), Pb(II), Fe(III), Co(II), and Cr(III) is proposed on Chelex-100 resin. The sample solutions at pH 6 were pumped through a micro-column at the rate of 2.0 mL/min by using a peristaltic pump. The elution of the retained analytes on the column was performed with 5 mL of 1 M HNO3 at 2.0 mL/min flow rate. The recoveries of analyte ions were greater than 95%. The relative standard deviations at the 50 mug/L levels of analyte ions were generally lower than 3% (N = 10). Recoveries of spike addition to bottled mineral water and artificial seawater were quantitative (>95%). The proposed preconcentration/separation method has been applied to determinations of the analyte contents of microwave and wet digested black tea and natural water samples. The validation of the developed preconcentration/ separation procedure was carried out by successfully analyzing standard stream sediment (GBW7310) reference material.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据