4.7 Article

Microfluidic chip-based liquid-liquid extraction and preconcentration using a subnanoliter-droplet trapping technique

向作者/读者索取更多资源

A robust and simple approach for microfluidic liquid - liquid (L - L) extraction at the subnanoliterscale was developed for on-chip sample pretreatment. Organic solvent droplets of a few hundred pL were trapped within micro recesses fabricated in the channel walls of a microfabricated glass chip. L - L extraction was performed by delivering aqueous samples through the channel, with the sample stream continuously flowing adjacent to the droplets. The analytes in aqueous streams were enriched within the droplet with high preconcentration factors owing to both phase transfer and dissolution of organic solvent into the bypassing aqueous sample. An aqueous solution of butyl rhodamine B (BRB) and 1-hexanol were used, respectively, as sample and extractant to demonstrate the performance of the system. The fluorescence intensity of the dye extracted into the droplet was monitored in situ by LIF. The system proved to be an efficient means for achieving high enrichment factors of over 1000, with sample consumption of a few mu L. Quantitative measurement of the extracted analyte was achieved with a linear response in the range 1 x 10(-9) - 8 x 10(-7) M BRB. The precision of the measured fluorescence values for a 10(-7) M BRB standard with a 12.5 min preconcentration period was 6.6% RSD (n = 5).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据