4.4 Article

Development of a multi-target screening analysis for 301 drugs using a QTrap liquid chromatography/tandem mass spectrometry system and automated library searching

期刊

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
卷 19, 期 10, 页码 1332-1338

出版社

JOHN WILEY & SONS LTD
DOI: 10.1002/rcm.1934

关键词

-

向作者/读者索取更多资源

A new multi-target screening (MTS) procedure for drugs in blood and urine for toxicological analysis has been developed using a hybrid triple-quadrupole linear ion trap mass spectrometer (QTrap) for the fast detection and identification of 301 forensically important drugs, e.g. tranquilizers (benzodiazepines), hypnotics, drugs of abuse (opiates, cocaine, amphetamines, cannabinoids), antidepressants, neuroleptics, and some cardiac drugs, in one single liquid chromatography/ tandem mass spectrometry (LC/MS/MS) analysis. Samples were extracted either with liquid-liquid extraction or solid-phase extraction. A multiple reaction monitoring (MRM) as survey scan and an enhanced product ion (EPI) scan as dependent scan were performed in an information-dependent acquisition (IDA) experiment. Finally, drug identification was carried out by library search with a newly developed MS/MS library based on EPI spectra at three different collision energies in positive mode. The advantage of this newly developed method is the possibility to detect and identify 301 drugs in one single LC/MS/MS run. Copyright (c) 2005 John Wiley & Sons, Ltd.