Tetraalkylammonium halides as chemical standards for positive electrospray ionization with ion mobility spectrometry/mass spectrometry

期刊

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
卷 19, 期 21, 页码 3051-3055

出版社

WILEY
DOI: 10.1002/rcm.2172

关键词

-

向作者/读者索取更多资源

Chemical standards for positive ion mode electrospray ionization ion mobility spectrometry/mass spectrometry (ESI(+)-IMS/MS) are suggested. The low clustering tendency of tetraalkylammonium halides makes them ideal chemical standards for ESI(+)-IMS/MS. A homologous series of these compounds forms a useful external standard for instrument testing and resolution calibration of an IMS instrument. Selected homologues or a mixture of tetraalkylammonium halides can be used as mobility standards in the analytical runs. Absolute and relative reduced mobilities were calculated for C-2-C-8, C-10 and C-12 tetraalkylammonium halides. Absolute reduced mobilities in nitrogen were 1.88, 1.56, 1.33, 1.15, 1.02, 0.92, 0.84, 0.73, and 0.67 cm(2) V-1 s(-1), for C-2-C-8, C-10 and C-12 tetraalkylammonium halides, respectively. Relative reduced mobilities (relative to 2,6-di-tert-butylpyridine) for the same compounds were 1.20, 1.00, 0.855, 0.743, 0.658, 0.59, 0.54, 0.47, and 0.43, respectively. Copyright (C) 2005 John Wiley & Sons, Ltd.