Dynamic shear stress in parallel-plate flow chambers

期刊

JOURNAL OF BIOMECHANICS
卷 38, 期 1, 页码 159-167

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.jbiomech.2004.03.020

关键词

fluid flow; dynamic shear stress; parallel-plate flow chamber; cells; dynamic loading

向作者/读者索取更多资源

An in vitro model using a parallel-plate fluid flow chamber is supposed to simulate in vivo fluid shear stresses on various cell types exposed to dynamic fluid flow in their physiological environment. The metabolic response of cells in vitro is associated with the wall shear stress. However, parallel-plate flow chambers have not been characterized for dynamic fluid flow experiments. We use a dimensionless ratio h/lambda(v), in determining the exact magnitude of the dynamic wall shear stress, with its oscillating components scaled by a shear factor T. It is shown that, in order to expose cells to predictable levels of dynamic fluid shear stress, two conditions have to be met: (1) h/lambda(v) < 2, where h is the distance between the plates and lambda(v) is the viscous penetration depth; and (2) f(o) < f(c)/m, where the critical frequency f(c) is the upper threshold for this flow regime, m is the highest harmonic mode of the flow, and f(o) is the fundamental frequency of fluid flow. (C) 2004 Elsevier Ltd. All rights reserved.