4.7 Article

Elimination efficiency of different reagents for the memory effect of mercury using ICP-MS

期刊

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/b511367a

关键词

-

向作者/读者索取更多资源

The present study was carried out to solve the problems of long washout time and non-linear calibration curves encountered in mercury analysis using inductively coupled plasma mass spectrometry. Comparisons of the washing efficiency for different reagents to eliminate the mercury memory effect were made. Of all the selected washing reagents, mercapto reagents such as 2-mercaptoethanol and L-cysteine could efficiently clean the instrument. Taking the toxicity and odor of mercaptoalkanol into account, L-cysteine was preferred as the most suitable washing reagent and was added into the standard and sample solutions. A good linear calibration curve was obtained with the correlation coefficient of 0.9999. The detetion limit for addition of L-cysteine was 0.024 mu g l(-1). Addition of 0.18% L-cysteine into the sample solution also facilitates the washout of mercury even using deionized water. The recoveries of mercury in certified reference materials, i.e. human hair and dogfish muscle, were 97.2% and 98.3%, respectively, when using 0.18% L-cysteine in the sample solutions.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据