Optimisation of dynamic reaction cell (DRC)-ICP-MS for the determination of Ca-42/Ca-43 and Ca-44/Ca-43 isotope ratios in human urine

期刊

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
卷 21, 期 3, 页码 297-304

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/b511741k

关键词

-

向作者/读者索取更多资源

A new method for the measurement of calcium isotope ratios (Ca-42/Ca-43 and Ca-44/Ca-43) in urine using DRC-ICP-MS after oxalate precipitation is presented. The isotope ratios were measured with a precision of approximately 0.15% RSD, limited only by counting statistics in samples diluted to a total calcium concentration of 3 - 4 mg L-1. The observed mass bias was substantial, approximately 5 - 10% per u. Careful mass bias correction using matrix and concentration matched solutions was essential and necessary in order to obtain accurate results. A methane gas flow of 0.6 ml min(-1) was used as a reaction gas in the dynamic reaction cell to remove polyatomic interferences (e.g., Ar-40(+) and (ArH2+)-Ar-40). Unfortunately, other interferences are formed in the dynamic reaction cell from secondary reactions with methane adducts. An interfering ion observed at m/z = 43 with a precursor ion of m/z = 17 was removed by applying an Rpq setting of 0.4, while an interfering ion on m/z = 48 with a precursor ion of m/z = 37 could not be removed. Despite the fact that the Ar-40 interference was successfully removed by the methane reaction gas, it was not possible to measure the Ca-44/Ca-40 and Ca-42/Ca-40 ratios with a precision only limited by counting statistics, most likely due to uncertainties from the detection system when measuring high count rates. The developed method was successfully applied for the calcium isotope ratio analysis of human urine samples enriched in Ca-42 and Ca-44. It was found that an atom enrichment of 1.3% was needed in the urine samples in order to obtain reliable results.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据